Section 3: Sweat Induction and Analysis

3.1 Performing the Sweat Test

NANODUCT
2019-01-31 12:03
Check Controls
Iontophoresis
Sweat Analysis
Recall Reading
WARNING!
Due to the remote possibility of ignition, never attempt iontophoresis on
a patient receiving oxygen-enriched respiratory therapy in an enclosed
space. With medical approval, remove the patient from that environment
during iontophoresis.
1 Assemble Equipment and Supplies
Make certain everything is on hand for the complete procedure:
• Nanoduct 1030
• Two holders
• Two holder straps with strap retainer posts
• Two Pilogel discs
• One sensor
• Electrode cable for connecting the iontophoretic electrodes or the sensor to
the induction/analysis module
A supply of deionized water, alcohol, cotton balls or gauze pads, cotton
swabs, powder-free gloves, and a roll of 1-inch wide plastic surgical tape (3M
Transpore™ recommended) is needed.
Make sure that the battery for the instrument is not dead, and that the gel discs
are rubbery, translucent and not cracked or otherwise damaged.
NOTE: The operator should wear powder-free gloves throughout the
iontophoresis and sweat analysis processes (graphics on the following pages do
not show the operator with gloves on; however, gloves should be worn as stated.)
2 Inspect Electrodes, Leads, and Cables
Clean the electrodes if necessary. Check the lead wires and insulation for
damage.
WARNING!
Replace the electrode cable assembly if wires, insulation or plastic housing
are cracked or frayed.
Connect the electrode cable assembly to the electrode socket.
3 Calibrate and Check Control Values
Turn the power on. Check for a low battery condition on the display. The arrow
on the display should indicate Check Controls on the menu. We recommend
checking control values (using the calibration plate mounted in the carrying case)
before each session. Refer to Section 4.3 to perform a calibration.
NOTE: In the SETUP Menu, selecting the Force Cal. option brings up the Check
Controls operation whenever the instrument is turned on.