Section 3: Sweat Induction and Analysis
NANODUCT
2019-04-15 12:03
Check Controls
Iontophoresis
Sweat Analysis
Recall Reading
Sweat Analysis
Suppress Readings
Display Reading
11 Insert Sensor Into Positive Holder
Insert the sensor (without being attached to the wiring harness) into the holder
(red locking ring) as described in steps 7a and 7b, taking care not to disturb the
holder or touch the bottom of the sensor.
Confirm that the instrument is ON.
12 Connect Sensor to the Red Sensor Connector
With the sensor in the holder, press the sensor connector fully onto the red
sensor, so that the two electrode pins seat in the mating receptacles (orientation
is not important).
NOTE: The sensor cable should be securely taped to the patient's arm
approximately 2 inches from the strain relief. Be sure to leave enough slack in the
cable on the sensor side of the tape-down to protect it from inadvertent tugging.
Make sure the tape-down doesn't disturb the sensor's even contact with the skin
surface. Make sure the black electrode cable is still securely taped down.
13 Activate Sweat Conductivity Analysis
A. On the screen, the pointer should be pointing to Sweat Analysis.
Immediately after the sensor is placed in the holder, press ENTER to begin
the sweat conductivity analysis.
B. The instrument prompts to display readings during testing or to suppress
the reading for recall at a later time. Press SELECT to make your selection
and press ENTER.
34