Section 4: Troubleshooting and Preventive Maintenance

Error Message Troubleshooting and Diagnosis

Error Message
Displayed
Low Battery
Fault
Calibration
Not Completed
Screen
Main Menu
Probable Causes:
Possible Solutions:
After replacing the batteries, if the same problem persists, contact
ELITechGroup for further instructions.
Check Controls
Probable Causes:
Possible Solutions:
If fault continues to occur, contact ELITechGroup for further
instructions.
Probable Causes / Possible Solutions
Battery voltage is below the low battery threshold.
Replace all four batteries.
An attempt to calibrate to a value other than 80 mmol/L.
Damaged electrode cable assembly or calibration plate.
Repeat calibration using the 80mmol/L position on the
calibration plate.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires.
43