Elitech NANODUCT 1030 User Manual page 46

Neonatal sweat analysis system
Table of Contents

Advertisement

Section 4: Troubleshooting and Preventive Maintenance
Error Message
Displayed
Initial Sweat Rate
Not Valid
Screen
Sweat Analysis
Probable Causes:
Recall Reading
Possible Solutions:
If fault continues to occur, contact ELITechGroup for further
instructions.
Probable Causes / Possible Solutions
Black electrode removed during sweat analysis.
Red sensor connector not connected to the sensor cell
before sweat flows into sensor cell.
Damaged electrode cable assembly.
Black electrode must remain attached during sweat
analysis for the Nanoduct to display a sweat rate value.
Connect red sensor connector to sensor cell immediately
after inserting sensor cell into the positive holder.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires. Replace
electrode cable assembly.
Repeat the sweat test one time.†