Section 3: Sweat Induction and Analysis
Wash and Dry Metal Plates
NANODUCT
2019-04-15 12:32
Check Controls
Iontophoresis
Sweat Analysis
Recall Reading
Recall Reading
2019-04-15 12:20
Mean Electrolyte
Concentration
61 mmol/L eq NaCl
Initial Sweat Rate
6.1
g/m
2
/min
ENTER To Continue
15 Remove Holders and Sensor, and Clean Electrodes
A. Disconnect the cable from the sensor. Remove the sensor from the holder
and discard the used sensor. Immediately and properly dispose of the
Pilogel disc from the negative (black) electrode.
WARNING!
Due to possible biological contamination, Nanoduct sensors are single use
only and must be discarded after use.
B. Disconnect the cable plug from the socket of the instrument and turn OFF
the power.
C. Remove the holders from the patient and gently wash and dry the entire
skin areas involved in the test.
D. Wash and dry the electrode plates.
16 Recall the Display Reading
If you selected "Suppress Readings," or if the control unit powers off, the last
reading can be recalled by selecting Recall Reading from the main menu and
pressing ENTER.
Recall Reading also outputs the last reading to the serial port where it can be
printed if a serial printer is connected to the port. The output includes the date
and time of the last reading and the reading in mmol/L eq NaCl.