Section 2 : Initial System Setup
Self Test Mode
ENTER To Continue
Connect Electrodes
To Check Current
#.### m.A.
Self Test Mode
ENTER To Continue
Connect To
Calibration Plate
To Check Operation
80 mmol/L eq NaCl
Self Test Mode
ENTER To Continue
Connect To Printer
SELECT Key
To Test Serial
Communication
3. Connect the electrodes by placing a Pilogel disc in each holder and then
holding the two discs together between two fingers to check the current.
The current should read approximately 0.5 mA. The instrument continues to
read and display the electrode current reading. Current that continually
decreases is an indication of low or weak batteries. Press ENTER to
continue with the Self Test.
4. Connect to the calibration plate to check for proper operation. The display
should read the value of the calibration point that the cable is connected to.
For example, if connected to the 80 mmol/L calibration point, then the
reading on the display should be 80 mmol/L.
5. Connect to the printer to test the serial communication. Press SELECT to
send data out the serial port. The date and time are sent out the port along
with the message "PRINT TEST" each time SELECT is pressed.
2019-04-21 15:24
PRINT TEST
Press ENTER to return to the SETUP MENU.
27