Section 4: Troubleshooting and Preventive Maintenance
Calibrating
Please Wait. . .
Calibration
Complete
2008-03-24 9:34
Check Controls
Electrolyte
Concentration
80 mmol/L eq NaCL
Exit
Calibrate
Check Controls
Electrolyte
Concentration
40 mmol/L eq NaCl
Exit
Calibrate
Check Controls
Electrolyte
Concentration
120 mmol/L eq NaCl
Exit
Calibrate
6. The Nanoduct calibrates to the 80 mmol/L (equivalent NaCl). The screen
indicates that the instrument is calibrating and to please wait. When the
calibration has been completed, the screen advances to the next screen.
7. When the calibration is completed, a Calibration Complete screen is
displayed for a few seconds. The time of the calibration also gets displayed.
Data is sent to the serial port.
8. Optionally for checking control values, connect the sensor cell connector to
any of the 3 control values (40, 60, 120). The value gets displayed as
shown below:
Select Exit and press ENTER to exit Check Controls.
51