Section 4: Troubleshooting and Preventive Maintenance
Error Message
Displayed
Fault
Sweat Analysis Was
Not Valid
*Reset the Nanoduct by doing the following: (1) Remove the two Phillips screws from the battery door
located on the bottom of the Nanoduct, then remove the door. (2) Remove one of the batteries from the
battery holder, wait for one minute, reposition the battery with correct polarity in the battery holder. (3)
Return the door to original position, insert the two Phillips screws and tighten. (4) Turn the Nanoduct on by
pressing the ON key.
†The sweat test should only be repeated one time within any 24-hour period. Consult a physician before
performing multiple tests on a patient within a 24-hour period.
‡The lithium coin cell backup battery used to power the clock is not accessible to the user and should only
be replaced by qualified service personnel.
NOTE: This instrument is designed and tested to meet IEC CISPR 22 / EN55022 and IEC/EN 60601-1-
2 standards for electromagnetic compatibility for medical electrical equipment. However, if you suspect
electromagnetic susceptibility or interference, reorient or relocate the equipment to correct the problem.
Screen
Sweat Analysis
Probable Causes:
Recall Reading
Possible Solutions:
Probable Causes / Possible Solutions
Insufficient sweat collected for analysis or air bubbles
present in sensor cell due to sensor cell not properly
latched with locking ring or holder not adequately secured
to patient.
Cable may have become detached from sensor cell or
Nanoduct.
Damaged electrode cable assembly.
Orifice on sensor cell blocked.
Ensure sensor cell is latched with locking ring and that
holder secured with adequate tightness to the patient's
limb. Do not overtighten and impede blood circulation.
Ensure that the cable is properly connected to both sensor
cell and Nanoduct.
Inspect the electrode cable assembly for any damage.
Do not use electrodes if there is any damage to the cable
insulation or if there are exposed bare wires. Replace
electrode cable assembly.
Repeat the sweat test one time using a new sensor cell.†
48