Setting Language; Self Test Mode - Elitech NANODUCT 1030 User Manual

Neonatal sweat analysis system
Table of Contents

Advertisement

Section 2 : Initial System Setup
CONFIGURE MENU
Exit
Set Clock
Set Options
Set Language
Set Language
English
French
German
Spanish
SETUP MENU
2019-01-31 12:03
Exit
Configure
Self Test
Demo Mode

Self Test Mode

ENTER To Continue
Battery Volt: x.xx
Setting Language
1. The display language is set from the CONFIGURE MENU. Select "Set
Language" and press ENTER. Language selections include English,
French, German, and Spanish.
2. Select the desired language and press ENTER. Control returns to the
CONFIGURE MENU with the selected language displayed.
Select "Exit" and press ENTER to return to the SETUP MENU.
Self Test Mode
The Self Test allows testing of current output, conductivity, and battery voltage.
1. From the SETUP MENU, select "Self Test" and press ENTER.
WARNING!
Do not run the Self Test while the module is connected to a patient.
2. Self Test Mode begins by displaying the battery voltage. A good battery
voltage reading should be greater than approximately 5.2 volts. Fresh
batteries should read approximately 6.4 volts.
26