Section 4: Troubleshooting and Preventive Maintenance

4.2 Cleaning the Instrument

When needed, the induction/analysis module surfaces and accessories should
be wiped down using a soft cloth dampened with mild detergent or 10%
household bleach solution.
Electrodes must be cleaned following each iontophoresis procedure as follows:
1. Remove any remaining gel from electrodes.
2. Use a cotton ball or swab moistened with purified water to clean each
electrode thoroughly.
3. If the electrode appears dirty after an extended idle period, or does not
clean with steps 1 and 2, use a small round piece of light duty cleaning pad
(such as 3M Scotch-Brite™ #7445) to buff the electrode surface.
CAUTION
Never use harsh abrasives such as steel wool, sandpaper, or emery cloth to
clean electrodes. Never scrape electrodes with metal tools. If the electrode
surface is scratched or pitted it does not perform as specified and must be 
replaced.
Use the following options to clean Nanoduct straps, holders and attachment strap
retainers or any other parts that come into contact with a patient:
Option A:
1. Soak the straps for 30 minutes in a freshly prepared 10% dilution of
household bleach.
2. Rinse soaked straps thoroughly in tap water.
3. Allow to air dry.
Option B:
1. Autoclave the straps for 30 minutes at 121 °C.
CAUTION
Do not autoclave the Nanoduct holders (red or black) or the attachment
strap retainer post. Autoclaving destroys these parts.
Option C:
1. Treat straps as disposables.
2. Discard straps after each use.
3. Purchase new straps from ELITechGroup.
49