Demo Mode - Elitech NANODUCT 1030 User Manual

Neonatal sweat analysis system
Table of Contents

Advertisement

Section 2 : Initial System Setup
SETUP MENU
2019-01-31 12:03
Exit
Configure
Self Test

Demo Mode

DEMO MODE
2019-01-31 12:03
Exit
Check Controls
Iontophoresis
Sweat Analysis
Recall Reading
Check Controls
Electrolyte
Concentration
80 mmol/L eq NaCl
Exit
Calibrate
Calibrating
Please Wait. . .
Calibration
Complete
2019-03-24 9:34
Demo Mode
The Demo Mode is used to simulate typical displays that are seen during
normal operation. This can be useful for demonstrating instrument function or to
familiarize the user with the instrument. No actual iontophoresis or sweat analysis
occurs in this mode and only simulated results display.
In Demo Mode, the instrument functions much like it would when running on
an actual patient. The display screens and menus appear as they would during
an actual test, with the exception that Demo Mode appears near the top of the
display. While running in this mode, the ENTER key acts as a shortcut key and
advances the instrument through the phases of the operation, allowing the next
display to be seen without waiting for the full instrument countdown.
The Check Controls option in Demo Mode works as it does under normal
operation. It is possible to check the readings using the calibration plate and
to also calibrate the instrument. When exiting Check Controls, the instrument
returns to the Demo Mode main menu.
28