DIX_ScopeRefresh
recent injection command to the associated DI Driver module. If multiple channels within the module are
operating, it is not possible to identify which channel is being displayed. Note that the resolution of the
current display is approximately 10 µsec sampling interval, and is for reference only. A box instrument
oscilloscope and current probe may be necessary for more detailed information. The voltage trace shown
is not recorded within injection commands. Therefore, the voltage trace will be constant with the last value
recorded before the injection command. Also, the current trace is not recorded outside of the injection
command and will be displayed as zero.
Module Temp
Tip Strip: Internal module temperature
Units: C
Detail: Indicates the internal DI Driver module temperature. The shutdown fault threshold is 80 C. If the
internal module temperature reaches 80 C, a
power supply and injector drive channels will be disabled until the fault is cleared.
Battery Voltage
Tip Strip: DI Driver battery voltage level
Units: V
is pressed, the graph will refresh with injector current, in Amps, during the most
DIX_ModuleTempFault
will be indicated and the internal boost
NI DIDS-2100 Series User Manual | © National Instruments | 67