Performance Test
Individual Test Description
2. Test Signal Level Accuracy Test
Description
This test checks the signal level accuracy of the E4981A source port output signal.
Specification
@23 ± 5 ⋅C
Signal Level [V]
0.1 to 1.0
Test equipment
Digital Multimeter
Interface Box
Test Leads (BNC), 2m
BNC(f)-BNC(f) Adapter
BNC(m)-BNC(m) Cable (122cm)
Dual Banana-BNC(f) Adapter
20
Spec.
± 5.0 %
Keysight 3458A
Keysight 04284-65007
Keysight 16048D
Keysight 1250-1830, 4ea
Keysight 8120-1840
Keysight 1251-2277
Chapter 2