Thermo Scientific ISQ EC Operating Manual page 56

Hide thumbs Also See for ISQ EC:
Table of Contents

Advertisement

5
Installation
Page 56
A common ground for all system modules will avoid ground loops, which can
create erratic results (for example, high baseline noise).
CAUTION—Electric Shock or Damage to the Device
Connecting the device to a line voltage higher or lower than
specified could result in personal injury or damage to the device.
Therefore, connect the device to the specified line voltage only.
Never connect the device to a power socket that is shared with other
equipment (for example, multiple sockets).
Do not use defective multiple socket or extensions cords, as they
could result in personal injury or damage to the device.
After the power to the device is turned off, the device is still
energized as long as the power cord is connected. Repair work on
the device while the device is connected to power could lead to
personal injury. Therefore, always unplug the power cord before
starting repair work inside the device. If you were instructed to
remove any covers or panels, do not connect the power cord to the
device while the cover or panels are removed.
ISQ EC and ISQ EM Mass Spectrometer Operating Manual