Emerson Rosemount 3490 Series Reference Manual page 27

4-20 ma + hart compatible controller
Hide thumbs Also See for Rosemount 3490 Series:
Table of Contents

Advertisement

Reference Manual
00809-0100-4841
Table 3-2: Connection Terminal Descriptions (Panel-mount Control Unit)
Terminal
1
2
3
4-6
7-9
10-12
13-15
16-18
19-21
22-24
25-27
28-29
30
(2)
31
(2)
32
(1)
Mains-powered control unit only
(2)
Direct current (dc) powered control unit only
Cabling between the Rosemount 3490 Series Control Unit ("control unit") and a
transmitter should be a screened (shielded), twisted-pair type with the cable screen
(shield) connected to terminal 3 (marked with earth/ground symbol) on the control unit.
The cable screen (shield) should be left unconnected at the transmitter end unless there is
a terminal specifically provided for this purpose. Cable runs should be separate from any
high voltage or mains cables to avoid crosstalk or interference.
Connect terminal 30 (intrinsically safe earth/ground) to a high integrity earth/ground
point if the transmitter connected to terminals 1 and 2 is in a hazardous area. This is a
special condition for some intrinsically safe approvals.
Related information
Product certifications
3.4.3
Power connections
When the control unit is powered by mains alternating current (ac) power, select the
voltage as 115 V or 230 V using the voltage-selector slide switch.
When the control unit is direct current (dc) powered, ensure the supply is adequate (15 to
30 Vdc). Do not exceed 30 Vdc.
A switch or circuit breaker should be installed in close proximity to the instrument, and
labeled as such. Although the Rosemount 3490 Series Control Unit meets all European
Emerson.com/Rosemount
Function
Loop supply
Current input
Cable screen Earth
RS232
Current output
Digital input 1 and 2
Relay 1
Relay 2
Relay 3
Relay 4
Relay 5
(1)
Mains input
I.S. Earth
Negative
Positive
Installation
January 2021
Terminal marking
24V
Iin
(Earth symbol)
RX-TX-0V
24V-Iout-0V
IN1-IN2-0V
NO-COM-NC
NO-COM-NC
NO-COM-NC
NO-COM-NC
NO-COM-NC
L-N
(Earth symbol)
-
+
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3491Rosemount 3492Rosemount 3493

Table of Contents