Restoring factory defaults

NOTE: Restoring factory defaults can also be carried out using the optional ADTi-100 and ADT View software.
See
www.renishaw.com/adt
for more information.
When realigning the readhead, reinstalling the system, or in the case of continued calibration failure, factory defaults should be restored.
To restore factory defaults:
Switch system off.
Obscure the readhead optical window or connect the 'Remote CAL' output pin to 0 V.
Power the readhead.
Remove the obstruction or, if using, the connection from the 'Remote CAL' output pin to 0 V.
The LED will start continuously flashing indicating factory defaults have been restored and the readhead is in installation mode (flashing set-up LED).
Check the readhead mounting / installation and recalibrate the system.
Switching Automatic Gain Control (AGC) on or off
NOTE: AGC can also be switched on or off using the optional ADTi-100 and ADT View software.
See
www.renishaw.com/adt
for more information.
The AGC is automatically enabled once the system has been calibrated (indicated by a Blue LED). AGC can be manually switched off by connecting the 'Remote CAL' output pin to 0 V for > 3 seconds < 10 seconds.
The LED will then be solid Green.
LED diagnostics
Mode
LED
Installation mode
Green flashing
Orange flashing
Red flashing
Calibration mode
Blue single flashing
Blue double flashing
Normal operation
Blue
Green
Red
Blank flash
Alarm
4 red flashes
NOTE: See
'Troubleshooting', page 25
for more information on diagnosing faults.
ATOM DX
linear installation guide
Status
Good set-up, maximise flash rate for optimum set-up
Poor set-up, adjust readhead to obtain Green flashing LED
Poor set-up, adjust readhead to obtain Green flashing LED
Calibrating incremental signals
Calibrating reference mark
AGC on, optimum set-up
AGC off, optimum set-up
Poor set-up; signal may be too low for reliable operation
Reference mark detected (visible indication at speed < 100 mm/s only)
Low signal or over signal; system in error