Section 1: Introduction
Button
Delete/Erase/Remove
Edit/Change User
Zero
Calibrate
Unselected
Selected
Table 9: Maintenance Function Keys
Button
Individual Prime Buttons
(A, B, C, D, E, F)
ABCDEF Prime button
Pattern Test
Volume Test
Line Flush
60-Sec Prime
QC/PM
Name
Deletes or erases the selected item.
Allows editing of an existing stain or cytocentrifuge
program. Allows manual entry of slide or specimen
information (stain or cytocentrifuge mode). Also allows
system administrator to edit user information.
Zeros the Level Detect sensors.
Calibrates the Level Detect system.
Shows an unselected option.
Shows a selected or enabled option.
Name
Primes the selected line.
Primes all (A, B, C, D, E, F) reagent lines simultaneously.
Performs Pattern Test to ensure nozzles are
clear of debris and spraying properly.
Performs Volume Test to verify the selected
nozzle volume is within the correct range.
Performs Line Flush function for A, B, and D reagent lines.
Runs the pumps for 1 minute to prime the lines.
Shows the Preventive Maintenance and Quality Control logs,
when enabled from the System Setup menu (Section 3.1).
Description
Description
21
21