Section 3: Controlling and Customizing Stainer Functions
2. Select Enable Reagent Tracking for each desired reagent (A, B, C, D, E, F).
3. Press Back to return to the System Setup menu.
4. Press Reagents to reveal the Reagent Information screen.
5. Press Change next to the desired Reagent (A, B, C, D, E, F).
6. Scan the barcode of each enabled reagent bottle.
7. Press Save.
8. Verify that the barcode appears on the Reagent Information menu.
9.
Repeat steps 4-8 for each reagent bottle that is enabled in QC Maintenance Tracking.
NOTE: Reagent Information can be accessed directly by pressing the bottle icons on the right side of the
Main menu, then reagent information can be scanned or manually entered by pressing Change next to the
desired reagent.