Calibrating The Touchscreen - Elitech AEROSPRAY CYTOLOGY 7522 User Manual

Slide stainer/cytocentrifuge
Table of Contents

Advertisement

Section 7: Solving Problems
NOTE: Calibration requires full, unopened (caps and seals in place) 500 mL bottles of reagent, placed in
the correct tray positions (due to different densities of each reagent type).
6. Place the correct reagent bottles in all enabled positions, and press Start.
NOTE: The calibration function ignores any disabled reagent line.
7. Press OK. Press Back twice to return to the Main menu.
8. Return the reagent bottles to the tray as indicated in Section 2.1 to prepare for staining.
NOTE: For accurate reagent level detection and calibration, dip tubes must follow their pre-formed coiled
shapes.

Calibrating the Touchscreen

1. Select and hold Standby/Ready for 5 seconds. A calibration screen with a target appears.
2. Select the center of the target with a finger, stylus, or similar tool. Another target will appear in a
different location.
3. Continue to press the center of the targets until all the targets have been pressed (five total). After
the fifth target is pressed, the instrument will save the touch screen calibration and return to the Main
menu.
77
77

Advertisement

Table of Contents
loading

Table of Contents