Section 3: Controlling and Customizing Stainer Functions
5. Select Change next to the appropriate reagent.
6. Scan the reagent bottle barcodes (Section 3.2) or manually enter the reagent information in the correct
fields.
7. Press Save.
8. Repeat steps 5-7 for each reagent.

Modifying Level Detect Functions

The Level Detect function alerts when the reagent is running low, or when the waste container is almost
full. Reagent and waste container monitoring can be turned On and Off from the System Setup screen. The
system defaults to On (Tray and External) for reagent monitoring and to Off for waste container monitoring.
See Section 2.2 for complete instructions.
1. From System Setup, select Level Detect.
2. Select the reagent monitoring options to be modified.
• To disable monitoring, select Off next to the appropriate reagents.
• To enable monitoring, select Tray next to the appropriate reagents.
• To monitor reagent F, select External.
• To monitor the waste container, select External.

Changing User Language

1. From System Setup, press Language.
2. Select the software language from the list on the left.