Section 2: Setup and Preparation for Operation
3. Press Level Detect to enter the Reagent Level Detect Setup menu:
NOTE: The stainer should be turned ON for at least 30 minutes before zeroing to stabilize level sensors.
The instrument can be used during this time.
4. Press Zero. The display shows:
5. Remove all reagent bottles and press Start. The display shows:
NOTE: Vibrations or bumps to the instrument or lab bench can cause inaccuracies in zeroing or calibration.
6. After zeroing, press OK. Press the Back button to exit to the System Setup menu.
7. Return the reagent bottles to their correct positions in the tray.
NOTE: For accurate reagent level detection and calibration, dip tubes must follow their pre-formed coiled
shapes.