Elitech AEROSPRAY CYTOLOGY 7522 User Manual page 10

Slide stainer/cytocentrifuge
Table of Contents

Advertisement

Section 1: Introduction
Category
Fuses
Ambient Temperature
Operating
Storage
Relative Humidity
Altitude
Pollution Degree
Heat Dissipation
Maximum
Average During Staining
Average While Idle
Maximum Sound Emission
Table 2: Performance Specifications
Category
Reagent Spray Nozzles
Reagents
NOTE: Use only ELITechGroup reagents. REF numbers
for this stainer include the following: SS-051A, SS-051B,
SS-051C, SS-051D, SS-051D2, SS-051E, and SS-051F.
See Appendix A for detailed information about reagents.
Stain Program Settings
Characteristics
2.0 Amp time-lag (Type T) (5 x 20 mm) (2
required). Manufactured by Littlefuse (Part No.
218002) or Bussman (Part No. GDC-2A)
15 to 30 °C (59 to 86 °F)
-10 to 50 °C (14 to 122 °F)
≤ 80% non-condensing
≤ 2000 m (≤ 6562 ft.)
2
512 Btu/hour
102 Btu/hour
41 Btu/hour
72 dB
Characteristics
Two reagent nozzles: ABCE (which controls
these reagents under microprocessor
control) and DF (hematoxylin)
A – SS-051A (EA-50 Stain)
B – SS-051B (Orange G Stain)
C – SS-051C (Bluing Agent)
D – SS-051D or SS-051D2 (Hematoxylin Stain)
E – SS-051E (Alcohol Wash)
F – SS-051F (Acid Alcohol Wash)
Two example programs: Progressive 1, Regressive 1
Number of specimen slides to stain: Selectable
from main screen – 4, 6, 8, 12, 16 or Full
Stain modes: Progressive or Regressive.
Intensity settings: 0-9 within each stain mode
Staining Customization: Each setting adjustable from 0-9
Nuclear/Htox (Hematoxylin)
Orange G
Bluing
Cytoplasmic/EA
Prewash Alcohol, Prewash Bluing
Acid Alcohol Wash, Bluing Wash,
Orange G Wash, Final Wash
End Spin
10
10

Advertisement

Table of Contents
loading

Table of Contents