Section 3: Controlling and Customizing Stainer Functions

Controlling Beeper Alerts

1. From System Setup, select Beeper.
2. Use the sliders to modify the beeper volume for Cycle Complete, Warnings, Errors, or Key Clicks.
3. Select Beep On Startup to turn the audible startup alert ON or OFF.

QC/Maintenance Tracking

Under system default settings, the following QC/Maintenance Tracking options are disabled:
• Enable Stain Slide Tracking
• Enable Cyto Slide Tracking
• Manual Entry
• Enable Preventive Maintenance Tracking
• Enable Reagent Tracking
Enable Stain Slide Tracking
To activate Stain Slide Tracking:
1. From System Setup, press QC/Maintenance Tracking.
2. Press Enable Stain Slide Tracking.
3. Press Back twice to return to the Main screen. Verify that the Start Button on the main screen reads
"Load Slides."
NOTE: Selecting Enable Stain Slide Tracking changes the Start button on the Main menu to "Load Slides."
38