4.4 Optimizing the Ion Source Parameters
ThermoStar™
Optimize the ion source parameters in regular intervals, in parti-
cular during the first time operation and after all service work on
the ion source.
The optimization procedure differs, depending on the existing
configuration and available test gas.
Optimization with test gas air → section 4.4.1
Optimization with calibration unit (PFTBA) → section 4.4.2
Optimization with test gas argon → section 4.4.3
4.4.1 ThermoStar™ with Test Gas Air
Optimize with air if no special test gas is available.
Start <TuneUp>
In the menu bar choose <Tune>
Choose <Ion Source>
Adjust all parameters as shown in the picture below
Choose <Display/Measure Data > for displaying the peak
groups of nitrogen and oxygen
Optimize:
Test at which RF Polarity the better peak shapes are
obtained and choose the corresponding polarity.
Set Field-Axis in such a way that a useful resolution
and peak shape are obtained.
Alternatingly optimize Extraction and Focus to
maximum peak height.
Quit:
Close <Tune Ion Source QMA 200>
Save the changes by confirming the prompt with <YES>
Close <TuneUp>
BG 805 071 BE
(9904)
GSD300.om
4.4.2 ThermoStar™ with Calibration Unit (PFTBA)
The calibration substance PerFluorTriButylAmin yields mass
peaks up to 614 amu. It is ideal for optimizing ion source settings
in the high mass range.
Start <TuneUp>
Turn off SEM/Emission:
In the menu bar choose <Setup>
Choose <SEM/Emission Control> (or ctrl+c)
Turn off SEM and Filament
Close the window with <OK>
Open the valve of the calibration unit:
In the menu bar choose <Manual>
Choose <DI/DO...>
Double click Valve to activate it
Close the window with <OK>
Wait 10s (compensation of the pressure rise caused
by opening the calibration unit)
Turn on SEM/Emission:
In the menu bar choose <Setup> (or ctrl+c)
Choose <SEM/Emission Control>
Turn on SEM and Filament
Close the window with <OK>
In the menu bar choose <Tune>
Choose <Ion Source>
Adjust all parameters as shown in the picture below
Choose <Display/Measure Data > for displaying the peak
group around mass 129 amu
Optimize:
Test at which RF Polarity the better peak shapes are
obtained and choose the corresponding polarity.
Set Field-Axis in such a way that a useful resolution
and peak shape are obtained.
Alternatingly optimize Extraction and Focus to
maximum peak height.