8 Troubleshooting

If the gas flow of the units is restricted or completely blocked,
check for the following causes:
2
Capillary clogged
27 and
2
Orifice clogged
34
If the fault is more severe, please contact the nearest
Pfeiffer Vacuum Service Center.

8.1 Shortening the Capillary

If the capillary tends to clog, it may be advisable to shorten it.

8.1.1 OmniStar™

Preconditions
Unit switched off
Capillary detached from measurement point
Length of capillary still adequate after it has been shortened
Required tools
Tube cutter (recommended →
or knife file
Bend the capillary
until it breaks
BG 805 071 BE
(9904)
GSD300.om
2
29
2
36)
Make two opposing
small indentations
Max. depth of
indentations 0.5 mm
Do not finish the
point of break
(capillary gets
clogged)
Turn on the unit and wait 10 min (turbo pump acceleration
time)
Check the pressure in the analysis chamber as follows:
Start Balzers Quadstar™ 422 <Service>
In the <Measure> menu choose <Total pressure>
The total pressure of the analysis chamber is displayed
Displayed pressure >4×10
Displayed pressure <4×10
Possible causes:
Capillary clogged at the end
⇒ Shorten the capillary by additional 10 mm
clogged
Orifice
⇒ Replace orifice (→
Capillary clogged across the entire length
⇒ Replace capillary (→
Connect the capillary to the measurement point
-6
mbar
-6
mbar
2
34)
2
29)
Troubleshooting
27