4.3 Optimizing the Ion Source Parameters
OmniStar™
Optimize the ion source parameters in regular intervals, in parti-
cular during the first time operation and after all service work on
the ion source.
The optimization procedure differs, depending on the existing
configuration, measurement task, and available test gas.
Optimizing with test gas air → section 4.3.1
Optimizing with test gas argon → section 4.3.2
4.3.1 OmniStar™ with Test Gas Air
Optimize with air if no special test gas is available.
Start the <TuneUp> program
Open the inlet valve:
In the menu bar chose <Manual>
Choose <DI/DO...>
To activate double click Valve
Close window with <OK>
In the menu bar choose <Tune>
Choose <Ion Source>
Adjust all parameters as shown in the picture below:
Choose <Display/Measure Data > for displaying the peak
groups of nitrogen and oxygen.
Optimize:
Test at which RF Polarity the better peak shapes are
obtained and choose the corresponding polarity.
Set Field-Axis in such a way that a useful resolution
and peak shape are obtained.
Alternatingly optimize Extraction and Focus to
maximum peak height
Quit:
Close <Tune Ion Source QMA 200>
Save the changes by confirming the prompt with <YES>
Close <TuneUp>
20
Initial start up
4.3.2 OmniStar™ with Test Gas Argon
Optimize with the test gas argon if argon is subsequently to be
analyzed. Setting of "Cathode" to 40 V prevents formation of Ar
ions and allows detection of low H
Make sure that argon is connected to the GSD 300 and
that the feeder lines are thoroughly purged.
Start the <TuneUp> program
Open the inlet valve:
In the menu bar chose <Manual>
Choose <DI/DO...>
To activate double click Valve
Close window with <OK>
In the menu bar choose <Tune>
Choose <Ion Source>
Adjust all parameters as shown in the picture below:
Choose <Display/Measure Data > for displaying the peak
groups of argon.
Optimize:
Test at which RF Polarity the better peak shapes are
obtained and choose the corresponding polarity.
Set Field-Axis in such a way that a useful resolution
and peak shape are obtained.
Alternatingly optimize Extraction and Focus to
maximum peak height.
Quit:
Close <Tune Ion Source QMA 200>
Save the changes by confirming the prompt with <YES>
Close <TuneUp>
O concentrations on mass 18.
2
BG 805 071 BE
(9904)
GSD300.om
++