Advanced Configuration Functionality
November 2019
Burst mode enables you to set the burst mode as OFF or ON:
OFF - Turns burst mode off; no data are broadcast over the loop
ON - Turns burst mode on; data selected under burst option are broadcast over the
loop
Request preambles
LOI menu path
Request preambles is the number of preambles required by the transmitter for HART
communications.
Response preambles
LOI menu path
Response preambles is the number of preambles sent by the transmitter in response to
any host request.
8.4

Configure LOI/Display

8.4.1
Flow and totalizer display
LOI menu path
Use flow display to configure the parameters that will appear on the LOI flowrate screen.
The flowrate screen displays two lines of information. Choose one of the following
options:
Flowrate, % of Span
Flow, Total A
% Span, Total A
Flow, Total B
% Span, Total B
Flow, Total C
% Span, Total C
8.4.2
Language
LOI menu path
Use language to configure the display language shown on the LOI. Choose one of the
following options:
84
Detailed Setup → Output Config → HART → Req Preams
Detailed Setup → Output Config → HART → Resp Preams
Detailed Setup → LOI Config → Flow Display
Detailed Setup → LOI Config → Language
®
Rosemount
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol
Reference Manual