Troubleshooting
November 2019
Table 12-16: Test D. Terminal to case ground
Test conditions
•
Location: uninstalled
•
Required equipment:
Multimeter
•
17 and safety ground
Table 12-17: Test E. Electrode to electrode shield
Test conditions
•
Location: uninstalled
•
Required equipment:
Multimeter
•
18 and 17
•
19 and 17
Table 12-18: Test F. Electrode shield to coil
Test conditions
•
Location: uninstalled
•
Required equipment:
Multimeter
•
17 and 1
12.9
Technical support and service
Email addresses:
Worldwide:
Asia-Pacific:
Middle East and Africa:
North and South America
United States
800-522-6277
Canada
+1 303-527-5200
Mexico
+41 (0) 41 7686
111
Argentina
+54 11 4837 7000 Germany
Brazil
+55 15 3238 3677 Italy
146
Expected
Potential cause
value
≤ 0.3 Ω
•
Moisture in terminal block
•
Leaky electrode
•
Process behind liner
Expected
Potential cause
value
∞Ω (<1 nS)
•
Shorted electrode
•
Leaky electrode
•
Moisture in terminal block
Expected
Potential cause
value
∞Ω (<1 nS)
•
Process in coil housing
•
Moisture in terminal block
flow.support@emerson.com
APflow.support@emerson.com
FlowTechnicalSupport@emerson.com
Europe and Middle East
U.K.
0870 240 1978
The Netherlands
+31 (0) 318 495
555
France
0800 917 901
0800 182 5347
8008 77334
®
Rosemount
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol
Reference Manual
00809-0100-8782
Corrective action
•
Clean terminal block
•
Replace terminal block
•
Replace sensor
Corrective action
•
Replace sensor
•
Clean terminal block
•
Replace terminal block
Corrective action
•
Replace sensor
•
Clean terminal block
•
Replace terminal block