Troubleshooting
November 2019
Table 12-16: Test D. Terminal to case ground
Test conditions
Location: uninstalled
Required equipment:
Multimeter
17 and safety ground
Table 12-17: Test E. Electrode to electrode shield
Test conditions
Location: uninstalled
Required equipment:
Multimeter
18 and 17
19 and 17
Table 12-18: Test F. Electrode shield to coil
Test conditions
Location: uninstalled
Required equipment:
Multimeter
17 and 1
12.9

Technical support and service

Email addresses:
Worldwide:
Asia-Pacific:
Middle East and Africa:
North and South America
United States
800-522-6277
Canada
+1 303-527-5200
Mexico
+41 (0) 41 7686
111
Argentina
+54 11 4837 7000 Germany
Brazil
+55 15 3238 3677 Italy
146
Expected
Potential cause
value
≤ 0.3 Ω
Moisture in terminal block
Leaky electrode
Process behind liner
Expected
Potential cause
value
∞Ω (<1 nS)
Shorted electrode
Leaky electrode
Moisture in terminal block
Expected
Potential cause
value
∞Ω (<1 nS)
Process in coil housing
Moisture in terminal block
flow.support@emerson.com
APflow.support@emerson.com
FlowTechnicalSupport@emerson.com
Europe and Middle East
U.K.
0870 240 1978
The Netherlands
+31 (0) 318 495
555
France
0800 917 901
0800 182 5347
8008 77334
®
Rosemount
8782 Slurry Magnetic Flow Meter Transmitter with HART Protocol
Reference Manual
00809-0100-8782
Corrective action
Clean terminal block
Replace terminal block
Replace sensor
Corrective action
Replace sensor
Clean terminal block
Replace terminal block
Corrective action
Replace sensor
Clean terminal block
Replace terminal block