9.14

Smart Meter Verification measurements

The Smart Meter Verification test will make measurements of the coil resistance, coil
inductance, and electrode resistance and compare these values to the values taken during
the sensor baseline process to determine the sensor calibration deviation, the coil circuit
health, and the electrode circuit health. In addition, the measurements taken by this test
can provide additional information when troubleshooting the meter.
Reference Manual
Parameter
Test Condition
Test Criteria
8785 Test Result
Simulated Velocity
Actual Velocity
Velocity Deviation
Xmtr Cal Test Result
Sensor Cal Deviation
Sensor Cal Test Result
Coil Circuit Test Result
Electrode Circuit Test Result
Parameter
Continuous Limit
Simulated Velocity
Actual Velocity
Velocity Deviation
Coil Signature
Sensor Cal Deviation
Coil Resistance
Electrode Resistance
mA Expected
mA Actual
mA Deviation
Advanced Diagnostics Configuration
November 2019
LOI menu path (Diagnostics →
Variables → MV Results → Manual
Results)
Test Condition
Test Criteria
MV Results
Sim Velocity
ActualVelocity
Flow Sim Dev
Xmtr Cal Verify
Sensor Cal Dev
Sensor Cal
Coil Circuit
Electrode Ckt
LOI menu path (Diagnostics →
Variables → MV Results → Continual
Res)
Test Criteria
Sim Velocity
ActualVelocity
Flow Sim Dev
Coil Inductnce
Sensor Cal Dev
Coil Resist
Electrode Res
4–20 mA Expect
4–20 mA Actual
AO FB Dev
103