Advanced Configuration Functionality
November 2019
Channel 2
Channel 2 is available as discrete output only.
Discrete output 2
LOI menu path
This parameter displays the configuration for channel 2.
Flow limit (1 and 2)
There are two configurable flow limits. Configure the parameters that will determine the
criteria for activation of a HART alert if the measured flow rate falls within a set of
configured criteria. This functionality can be used for operating simple batching
operations or generating alerts when certain flow conditions are met. This parameter can
be configured as a discrete output if the transmitter was ordered with the auxiliary output
suite (option code AX) and the outputs are enabled. If a discrete output is configured for
flow limit, the discrete output will activate when the conditions defined under mode
configuration are met. See
Control
LOI menu path
This parameter turns the flow limit HART alert ON or OFF.
ON The transmitter will generate a HART alert when the defined conditions are met. If a
OFF The transmitter will not generate an alert for the flow limit.
78
Mode
discrete output is configured for flow limit, the discrete output will activate when
the conditions for mode are met.
Rosemount
Detailed Setup → Output Config → DI/DO Config → DO 2
below.
Flow 1: Detailed Setup → Output Config → DI/DO Config →
Flow Limit 1 → Control 1
Flow 2: Detailed Setup → Output Config → DI/DO Config →
Flow Limit 2 → Control 2
®