Optimising The Static Alignment; Checking The Static Alignment; Optimising During Dynamic Trial Fitting; Resetting To Factory Defaults - Otto Bock 3R120 Instructions For Use Manual

Puro spring
Table of Contents

Advertisement

Available languages

Available languages

6.3 Optimising the static alignment

6.3.1 Checking the static alignment

► To check the static alignment, position the patient on the measuring device as shown in
26
Fig. 
on page 19.
6.3.2 Optimising the static alignment
► In the sagittal plane, check the course of the load line 
2
point 
and correct this as shown if needed (see Fig. 
► In the frontal plane, check the course of the load line 
3
thetic foot,
 prosthetic knee joint,
28
(see Fig. 
on page 20).

6.4 Optimising during dynamic trial fitting

6.4.1 Resetting to factory defaults

► Before performing the exercises in the sections that follow, reset settings that have been
changed to factory defaults according to Fig. 

6.4.2 Practicing sitting down and standing up

CAUTION
Adjusting the Settings
Fall due to incorrect or unfamiliar settings
Only adapt the settings to the patient gradually.
Explain the effects of the adjustments on the use of the prosthesis to the patient.
► The patient has to learn safe operation of the lock for all exercises (lock by loading at full
extension/unlock using the lock slide).
► CAUTION! With the patient, practice sitting down 
page 21).
► CAUTION! With the patient, practice standing up 
up, it is necessary to wait until the joint moves to extension again.

6.4.3 Practicing walking

► CAUTION! Using parallel bars or a handrail 
31
Fig. 
on page 22).
► NOTICE! Be sure to establish functional exterior foot rotation 
► Check the exterior rotation 
► Check socket flexion 
► Check socket adduction 
► During the dynamic trial fitting, check to see if the deviations listed in figure 
occur and correct them correspondingly:
1
 The resistance to movement for extension during the swing phase is too large
2
 During the extension in the swing phase, there is no resistance to movement, or there is
but it is too small
3
 The resistance to movement for flexion during the swing phase is too large
4
 The resistance to movement for flexion during the swing phase is too small
5
 The resistance to movement for flexion during the swing phase is too large due to the
attached stops for flexion damping
6
 The resistance to movement for flexion during the swing phase is too small due to the
detached stops for flexion damping
4
 Spina iliaca anterior superior) and correct if needed
4
of the prosthetic knee joint.
5
.
6
.
1
relative to the alignment reference
27
on page 19).
1
relative to the marked points (
29
on page 20.
1
and standing up 
3
with a highly flexed joint. While standing
1
, practice walking 
3
.
2
 pros­
30
2
(see Fig. 
2
with the patient (see
32
on page 23
Ottobock | 47
on

Advertisement

Chapters

Table of Contents
loading

This manual is also suitable for:

3r120 kd3r120 hdPuro spring 3r120Puro spring 3r120-kdPuro spring 3r120-hd

Table of Contents