Rack Mounting the N432A
Rack mounting information
Disconnect device requirements
During rack mounting, a disconnect device, for example, an IEC320 appliance
coupler, must be used. Consideration must be given to the following requirements
when installing the N432A in a rack:
– An adequately rated switch (250 V, 10 A) or a circuit breaker (250 V,
– It should be in close proximity to the N432A and within easy reach of the
– It shall be marked as a disconnect device.
Ventilation requirements
When installing the N432A in a cabinet, the convection into and out of the N432A
must not be restricted. If the total power dissipated in the cabinet is greater than
800 W, then forced convection must be used.
Keysight N432A Installation Guide
5 A) shall be included in the rack.
operator.