Final Calibration For V11.10 And Later - Fluke ScopeMeter 190-202 Service Manual

Hide thumbs Also See for ScopeMeter 190-202:
Table of Contents

Advertisement

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

Final Calibration For v11.10 and later

Before you start the final calibration, do the Warming-Up and Pre-Calibration.
The final calibration requires input conditions that are described in each step.
After a step starts, steps that require the same input conditions are done
automatically. For example, if you start calibration step CL 0850, the calibration
can include step CL 0869 and at the end the display shows CL 0799: READY.
You must always start the Final Calibration at the first step. See the
Warming-Up 2, Warm-Up Final, and ADC Timing section. Starting at
another step will make the calibration invalid.
If you do calibration step N (for example, step CL 0581), then return to a previous
step (for example, step CL 0580), and then calibrate this step, the complete final
calibration becomes invalid; then you must repeat the calibration starting at the
Warming-Up 2, Warm-Up Final, and ADC Timing section.
It is allowed to repeat a step that shows the status :READY by pressing
again.
Error messages
Proceed as follows if an error message
during calibration:
• if
nnnn
the current step.
• if
nnnn
calibration in the Warming-Up 2, Warm-Up Final, and ADC Timing section.
If the error persists, the Test Tool is defective.
5-10
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Note
<5000, check input signal and test leads and press
5000, check input signal and test leads and repeat the final
shows on the display
ERROR: nnnn
F2
F3
to repeat