Measure Input Resistance and Capacitance
You can push the MODE softkey to move through the functions in the sequence shown,
or you can push to return directly to the OTHER modes menu.
Each option in the Impedance/Capacitance (MEAS Z) menu is given below.
•
Measured @ SCOPE terminal Shows the location of the measured input.
•
MEASURE Shows the type of test. You can select res 50 Ω or res 1 MΩ
termination (for impedance) or cap (capacitance).
•
MODE Shows the Calibrator is in MEAS Z mode. Use the softkey to change modes
and open the menus for the other oscilloscope calibration modes.
If you have selected Capacitance measurement, the menu shows as follow:
•
SET OFFSET With the cable disconnected at the oscilloscope but still connected at
the Calibrator, push SET OFFSET to cancel the capacitance of the Calibrator and
cable. Push again (CLEAR OFFSET) to cancel the offset.
Input Impedance Measurement
With MEAS Z mode selected, do the subsequent procedure to measure the input
impedance of an oscilloscope:
1. Use the MEASURE softkey to select "res 50Ω" or "res 1 MΩ" termination.
2. Connect the SCOPE terminal on the Calibrator to Channel 1 on the oscilloscope.
3. Push to initiate the measurement.
Input Capacitance Measurement
With MEAS Z mode selected, do the subsequent procedure to measure the input
capacitance of an oscilloscope:
SC600 Oscilloscope Calibration Option
Measure Input Resistance and Capacitance
9
gjh064.eps
gjh065.eps
9-27