717 Specifications; 718 And 718Ex Specifications - Fluke 71 Series Calibration Manual

Process calibrators
Hide thumbs Also See for 71 Series:
Table of Contents

Advertisement

718 and 718Ex Specifications

Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C.
To display firmware version, start with the unit off, press and hold M, then press O.
Model
Range SI
718-1G
(-1 to 1) PSI
718-30G
(-12 to 30) PSI
718-100G
(-12 to 100) PSI
718-300G
(-12 to 300) PSI
Accuracy: Pressure Accuracy is +/- 0.05% of positive range
Temperature coefficient: .01 % of range per °C for temperature ranges of between −10 °C to 18 °C
and 28 °C to 55 °C.
Range
Range
718: Version 2.9
and lower
24 mA
718: Version 3.0
24 mA
and higher
718 EX
24 mA
Table 23. Pressure Specifications
For version 1.3 or higher
Range Metric
(-7 to 7) kPa
(-83 to 207) kPa
(-83 to 690) kPa
(-83 to 2068) kPa or 2.1 mPa
Table 24. Pressure Display, Pressure Module Input
Resolution
Refer to the Instruction Sheet for the pressure module
Table 25. DC mA Input
Resolution
0.001 mA
0.001 mA
0.001 mA
Max SI
5 PSI
60 PSI
200 PSI
375 PSI
Accuracy, ± ( % of Reading + Counts)
0.025 + 1
0.015 + 2
0.02 + 2
Process Calibrators
Specifications
Max Metric
34.5 kPa
413 kPa
1379 kPa or 1.4 mPa
2586 kPa or 2.6 mPa
Accuracy
17