Instructions for Use

Instructions for Use

Introduction

Align the Phoroptor Head with the patient to ensure you are taking reliable measurements. Once the
Phoroptor Head is powered on and completes an initialization process, you can then align it and take
measurements.

Aligning the Phoroptor Head

Initial Alignment
The first alignment is the general alignment of the Phoroptor
Head to the patient.
1. Place the back side of the Phoroptor Head in front of
the patient, centering the head in front of the patient's
eyes. (Please refer to Figures IU-01 and IU-02.) The
Forehead Rest is an applied part that contacts with
the patient. Position the patient's head against the
forehead rest.
2. Use the Illuminated Bubble on the Phoroptor Head to
ensure the Phoroptor Head is level. (Please refer to
Figure IU-04.) If the Phoroptor Head is not level, refer
to the Setup - Leveling the Phoroptor Head section of
this manual.
3. Once the Phoroptor Head is level, physically adjust the
Phoroptor Head to line up with the patient's left eye.
(Please refer to Figure IU-03.) Do not worry about
aligning the right side of the Phoroptor Head with the
patient's other eye. This is adjusted later.
Figure IU-04, Level
52
Level Bubble
Forehead
Rest
Figure IU-01, Back Side of
Phoroptor Head
Figure IU-02, Placement
Figure IU-03, Adjust Phoroptor Head
16241-101 Rev. D