Cleaning and Maintenance
Cleaning and Maintenance
Forehead Rest Cleaning and Disinfection
For hygienic reasons, you can clean the Forehead Rest with a clean cloth moistened with a mild
detergent solution (1 cc of liquid dish soap to one liter of clean, filtered water (filtered below 5 microns)).
Note: If the Forehead Rest must be sanitized, you can use a 70% isopropyl alcohol wipe. Be sure the
Forehead Rest is completely dry before positioning another patient.
Fuse Replacement
1. Unplug the Central Unit power cord.
2. Squeeze in the tabs on the Fuse Holder and pull the Fuse Holder out.
Note: Tweezers or a similar tool can be used to remove the fuse holder.
3. Remove the fuses that require replacement, and install new fuses.
Note: Refer to the Specifications section of this manual for fuse specifications.
4. Install the Fuse Holder into the Central Unit until it snaps into place.
Battery Replacement
Time and Date are maintained in memory by a long-life battery (CR 2032 Lithium Coin Cell Battery),
in the Controller. This long-life battery lasts many years. If the Time and Date do not work, contact
Reichert (Please refer to the Troubleshooting section of this manual).