Guidance Tables
Table 202 – Guidance and Manufacturer's Declaration
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Phoroptor VRx is suitable for use in electromagnetic environment specified below. The
customer or user of the Phoroptor VRx should ensure that it is used in such an environment.
Immunity
Test
ESD
IEC 61000-4-2
EFT
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage
Dips/Dropout
IEC 61000-4-11
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
16241-101 Rev. D
(continued)
Electromagnetic Immunity
All Equipment and Systems
IEC 60601
Compliance
Test Level
Level
±6kV Contact
±6kV Contact
±8kV Air
±8kV Air
±2kV Mains
±2kV Mains
±1kV I/Os
±1kV I/Os
±1kV Differential
±1kV Differential
±2kV Common
±2kV Common
>95% Dip
>95% Dip
for 0.5 Cycle
for 0.5 Cycle
60% Dip
60% Dip
for 5 Cycles
for 5 Cycles
30% Dip
30% Dip
for 25 Cycles
for 25 Cycles
>95% Dip
>95% Dip
for 5 Seconds
for 5 Seconds
3A/m
3A/m
Electromagnetic
Environment
- Guidance
Floors should be wood, concrete, or ceramic
tile. If floors are synthetic, the R/H should be at
least 30%.
Mains power quality should be that of a
typical residential, commercial, or hospital
environment.
Mains power quality should be that of a
typical residential, commercial, or hospital
environment.
Mains power quality should be that of a
typical residential, commercial, or hospital
environment. If the user of the Phoroptor
VRx requires continued operation during
power mains interruptions, it is recommended
that the Phoroptor VRx be powered from an
uninterruptible power supply or battery.
Power frequency magnetic fields should be that
of a typical residential, commercial, or hospital
environment.
107