Instructions for Use
Cross Cylinder - Smart Test
Axis Adjustment
1. Touch XCyl on the menu bar at the bottom of the main screen.
2. Touch SMART to begin the Smart Cylinder test.
•
The Smart Cross Cylinder test is performed monocularly.
•
Spherical equivalent is maintained during the Cylinder power adjustment.
If no Cylinder power is present, the Astigmatism check is done first, followed by the Axis
•
adjustment.
•
If default charts are selected for the XCyl test, these charts are presented to the patient.
Otherwise, select a chart from the available charts section.
3. Turn the CONTROL KNOB to change the position of the conventional Cross Cylinder.
4. Ask the patient: "Which is better (sharper and clearer), 1 or 2?"
5. Change the Cylinder Axis in the direction the patient prefers.
6. Press the CONTROL KNOB, or touch the NUMBER (1 or 2), to enter the patient's preferred
selection.
Note: Touch Test to switch the type of XCyl test (Manual/Smart/Split Cyl).