Mounting Mag 1100 F - Siemens SITRANS MAG 1100 Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS MAG 1100:
Table of Contents

Advertisement

Vibrations
4.6

Mounting MAG 1100 F

www.eltra-trade.com
Avoid strong vibrations.
Figure 4-7
Avoid vibrations
CAUTION
In applications with strong vibrations, Siemens recommends remote mounting of the
transmitter!
The MAG 1100 F sensor has an integrated clamp connection and is designed for installation
between two adapters, which are supplied separately. The adapters are available for a variety
of standards according to: ISO, DIN, SMS, BS and DS, for direct welding into dairy pipes or
with clamp- or threaded fittings.
At assembly, a gasket is located in the recess in the face of the adapter, and the adapter is
then fastened to the MAG 1100 F sensor with a clamp ring.
Sensor
Gasket
Process connection adapter
Clamp ring
info@eltra-trade.com
Installing/Mounting
4.6 Mounting MAG 1100 F
27
+421 552 601 099

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents