Connecting
5.2 Remote installation
1. Unscrew and remove terminal box lid.
2. Mount the two terminal blocks as shown and insert electrode cable plug
3. Remove SENSORPROM
4. Fit the ½" NPT or M20 cable glands for supply and output cables.
36
www.eltra-trade.com
0 and 83) and coil cable plug
Terminals electrode cable: 82, 0 and 83
Terminals special electrode cable: 84, 83, 0, 82 and 81
Terminals coil cable: 85; 86
Note
Special electrode cable
Special electrode cable must be connected to terminals 84, 83, 0, 82 and 81.
Figure 5-1
Terminal box without blocks
®
see relevant transmitter operating instructions.
(terminals 85 and 86).
unit from sensor and mount it on connection plate in transmitter,
info@eltra-trade.com
(terminals 82,