Troubleshooting; Improving The Application; Sensor Check - Siemens SITRANS MAG 1100 Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS MAG 1100:
Table of Contents

Advertisement

Troubleshooting

7.1

Improving the application

Resolving problems
7.2

Sensor check

Requirement
Sensor check
www.eltra-trade.com
The first step in the troubleshooting procedure is to check for some easily resolved problems.
Check that:
● Sensor and SENSORPROM unit correspond (serial numbers)
● The sensor is properly installed.
● The sensor is located in a vibration-free position. Vibrations can disturb the sensor and
therefore cause measurement error.
● The sensor is filled with liquid and liquid only. Air or gas bubbles in the liquid cause instability
and can result in measurement errors.
Note
The liquid must be homogeneous in order to achieve high-accuracy measurements. If the
liquid contains solid particles of greater density than the liquid, then these solids can
precipitate, especially at low flow rates, which will cause instability in the sensor and lead
to measurement errors.
1. Make sure that the serial numbers on the sensor and the SENSORPROM
2. Make sure that the sensor is installed as described in the installation chapter (Page 19) of
the sensor Operating Instructions.
3. Flush the pipe systems and the sensor for several minutes at maximum flow rate to remove
any air bubbles which may be present.
To check the SITRANS F M sensors the following test instruments will be required:
● Digital Meter/Multimeter
● Megger
● (Moving Coil Meter)
Remove the transmitter from the sensor or remote position before making the following checks.