Installation In Partially Filled Pipes - Siemens SITRANS MAG 1100 Operating Instructions Manual

Electromagnetic flowmeters
Hide thumbs Also See for SITRANS MAG 1100:
Table of Contents

Advertisement

4.3

Installation in partially filled pipes

Installation in partially filled pipes
Installation in large pipes
www.eltra-trade.com
Note
Empty pipe detection
For applications with empty pipe detection, the sensor can be tilted 45°, as shown above.
For partially filled pipes or pipes with downward flow and free outlet the flowmeter should be
located in a U-tube.
The flowmeter can be installed between two reducers (for example DIN 28545). At α ≤ 8° the
following pressure drop curves apply. The curves are applicable to water.
info@eltra-trade.com
Installing/Mounting
4.3 Installation in partially filled pipes
+421 552 601 099
21