Reflection Measurements - Keysight N9340B Demo Manual

Handheld spectrum analyzer
Table of Contents

Advertisement

Section 9

Reflection measurements

Figure 9. Reflection measurement setup
This demo shows a return loss measurement on a 300 MHz bandpass fi lter.
You will need a directional bridge such as the Keysight 86205A to perform this
measurement. Before making the measurement, calibrate the test setup using
a calibration standard with a known reflection coefficient.
The calibration standard for refl ection measurements is usually a short circuit
connected at the reference plane (the point at which the DUT is connected).
A short circuit has a refl ection coeffi cient of 1 (0 dB return loss); it refl ects
all incident power and provides a convenient 0 dB reference.
Short
circuit
or
DUT
If possible, use a coupler or bridge with the correct test port connector for
both calibrating and measuring. Any adapter between the test port and DUT
degrades coupler/bridge directivity and system source match. Ideally, you
should use the same adapter for the calibration and the measurement. Be sure
to terminate the second port of a two port device.
Please refer to the following procedures when connecting the DUT to the
instrument:
1.
Connect the DUT to the directional bridge as shown in Figure 9 and
terminate
the unconnected port of the DUT.
2.
Connect the RF OUT port of the analyzer to the directional bridge.
3.
Connect the RF IN port to the coupled port of the directional bridge.
CAUTION:
Excessive signal input may damage the DUT. Do not exceed the maximum power
that the DUT can tolerate.
17
Coupled port
N9340B
100 kHz - 3.0 GHz
HANDHELD SPECTRUM ANALYZER
PRESET
BW/
SYS
M ODE
M EAS
TRACE
AM PTD
SPAN
SWP
1
2
3
ABC
DEF
GHI
4
5
6
JKL
M NO
PQR
7
8
9
STU
VWX
YZ_
0
SAVE
LIM IT
M ARKER
ESC/ CLR
ENTER
FREQ