Keysight N9340B Demo Manual page 24

Handheld spectrum analyzer
Table of Contents

Advertisement

Section 12
AM/FM, ASK/FSK Demodulation (continued)
Figure 15. AM demodulation
AM demodulation analysis
Instruction
Connect the signal generator to the RF IN
connector and turn on the signal generator's
AM and RF output.
Set the signal generator to 100 MHz, AM; AM
depth, 50%; amplitude, –15 dBM; modulation
rate, 1 kHz.
Turn on the AM demodulation analysis
function.
Set the carrier frequency of the source.
Set the attenuation state to auto.
Set the Y scale.
Set the AM depth limit and turn on the limit
function.
Save the result.
Detailed metrics provide you with a complete understanding of the AM signal.
24
Key strokes
[MODE], rotate the knob to highlight
Demodulation Analysis, [ENTER], {AM}
{Carrier Freq}, [100], {MHz}
{More 1 of 3}, {More 2 of 3}, {Attenuation
{Auto}}, {More 3 of 3}
{Y Scale}, {Auto Scale}, {Return}
{More 1 of 3}, {Limits}, {AM Depth UP},
[51], {%}, {AM Depth Low}, [49], {%},
{Limits on)}, {Return}
[SYS], {File}, enter the file using the
numeric keypad, {Save}