Automation (continued)
Remote control the N9340B via LAN.
Instruction
Connect the equipment as shown in Figure 22.
Next, power on the equipment.
Set the N9340B IP address with DHCP mode.
You can also set the N9340B IP address
with static mode such as a static IP address:
192.168.0.112.
Run the Keysight connection expert to
automatically find the N9340B.
Note: After clicking Add Instrument, if the
Add Instrument dialog box comes up, click
the interface Add LAN instrument on LAN
(TCPIP0).
You can manually add the N9340B into the
Keysight connection expert.
Run the N9340 PC software on the PC.
Note: Once the N9340B enters into remote
control mode, you can set up the N9340B's
parameters, and transfer files between the PC
and the N9340B with the File Manager and
monitor of the N9340 PC software.
32
Key strokes
[SYS], {Setting}, {IP Admin}, {IP Address
{DHCP}}
[SYS], {Setting}, {IP Admin}, {IP Address
{Static}}, [192.168.0.112], {ENTER}
Run the Keysight connection expert. The
connection expert is a program in the
Keysight IO libraries.
Click Refresh All.
Click the Add Instrument shortcut key on
the toolbar.
Click the Auto Find Discover Local
Instruments shortcut key on the toolbar.
Select the item corresponding to the IP
address of the N9340B.
Click Ok.
Click the Add Instrument shortcut key on
the toolbar.
Click the Add Address shortcut key or
manually add a known IP address or
hostname.
Select Use IP Address item and input the
IP address of the N9340B.
Select *IDN query, click Test Connection,
and the prompt "this instrument is
present" message will appear.
Click Ok.
Click the N9340B PC software shortcut
key on the PC screen.
Click the Monitor shortcut key in the
toolbar.
Highlight the corresponding IP address
and click ENTER in the Instrument list
dialog box. The monitor panel will open.