Measure Automatically

Click
to call up the menu panel. Click the Measure softkey on panel to
display the Measure menu. At most 8 types of measurements could be
displayed on the bottom left of the screen.
The oscilloscopes provide 39 parameters for auto measurement, including
Period, Frequency, Mean, PK-PK, RMS, Max, Min, Top, Base, Amplitude,
Overshoot, Preshoot, Rise Time, Fall Time, +PulseWidth, -PulseWidth,
+Duty Cycle, -Duty Cycle, Screen Duty, FRR, FRF, FFR, FFF, LRR, LRF,
LFR, LFF, Delay A→B , Delay A→B , Cycle RMS, Cursor RMS, FRR,
FRF, FFR, FFF, LRR, LRF, LFR, LFF, Phase A→B
+PulseCount, -PulseCount, RiseEdgeCnt, FallEdgeCnt, Area, and Cycle
Area.
The "Automatic Measurements" menu is described as the following table:
Function Menu
Add
Remove
Snapshot
Source
Gating
Set
Statistics
4.Advanced User Guidebook
Description
Meas Type
Select the measure types
(left menu)
Source
Select the source
CH1 CH2
Add
Add
(shown at the left bottom, you could
only add 8 types at most)
Select the types need to be deleted.
Meas Type
The selected type and source are
(left menu)
shown in the Remove menu on the
right.
Remove
Remove the selected measure type
Remove All
Remove all the measures
Show all the measures of the snapshot
ON
source
OFF
Turn off the snapshot
CH1
Select the snapshot source
CH2
Screen
Select "Screen" or "Cursor" to set the c
Cursor
ursor range
ON
select "ON" or "OFF" for measurement
OFF
value statistics
Reset
Restart statistics on statistics