Use Executive Buttons - Owon TAO3000 Dual-Channel Series User Manual

Tablet oscilloscopes
Hide thumbs Also See for TAO3000 Dual-Channel Series:
Table of Contents

Advertisement

menu. Set Horizontal tolerance and Vertical tolerance in the right menu.
Select Create in the right menu to create the rule.
3. Set output type: Select Output in the bottom menu to enter output option
setting. Choose any one or two of the options "Pass", "Fail" or "Beep".
"Pass" and "Fail" are mutually exclusive options, which could not be
chosen simultaneously. "Stop" means stop once the condition satisfies your
setting.
4. Begin to test: Select Operate in the bottom menu, switch it to Start, the
test will begin.
5. Save rule: Select SaveRule in the bottom menu. Select the save location
in the left menu, and then select Save in the right menu to save the rules,
which could be called up at once when need. Select Load to call up the rule
saved.
Note:
1. When Pass/Fail is ON, if XY or FFT is ready to run, then Pass/Fail will be
closed; under the mode of XY or FFT, Pass/Fail is unable.
2. Under the mode of Factory, Auto Scale and Auto Set, Pass/Fail will be
closed.
3. When no save setting left in the rule save, tip will be given to show "NO
RULE SAVED".
4. Under the status of stop, data comparing will stop, and when it goes on
running, the number of Pass/Fail will increase from the former number, not
from zero.
5. When the waveform playback mode is on, Pass/Fail is used to test the
played-back waveform specially.

Use Executive Buttons

Executive Buttons include Copy, Default, Run/Stop, Single, Autoset.
Copy
You can save the waveform by just pushing the Copy panel button in any user
interface. The source wave and the storage location are according to the
settings of the Save function menu when the Type is Wave. For more details,
please see "Save Function Menu" on P55.
Default
Shortcut keys to restore factory settings.
Run/Stop
Enable or disable sampling on input signals.
Notice: When there is no sampling at STOP state, the vertical division and the
horizontal time base of the waveform still can be adjusted within a certain
range, in other words, the signal can be expanded in the horizontal or vertical
direction.
4.Advanced User Guidebook