 Trigger control
 FFT
Operation steps:
1. Click
to call up the menu panel. Click the XY softkey on panel to
switch to the ON state.
2. To make the XY view full screen, select Full Screen as ON in the bottom
menu.

Use Pass/Fail function

The Pass/Fail function monitors changes of signals and output pass or fail
signals by comparing the input signal that is within the pre-defined mask.
Click
to call up the menu panel. Click the P/F softkey on panel to switch to
the ON state.
The Pass/Fail menu:
Function
Setting
Menu
Start
Operate
Stop
Pass
Fail
Output
Beep
Stop
Info
Source
Horizontal
Rule
Vertical
Create
Number
SaveRule
Save
Load
The description of Pass/fail Menu is shown as the follows:
Pass/Fail test:
Detect whether the input signal is within the limits of the rule, if it exceeds limits
of the rule, it is "Fail"; otherwise it is "Pass". Also, it can output fail or pass
signal by built-in and configurable output port. To run the test, read the
following steps:
1. Click
to call up the menu panel. Click the P/F softkey on panel to switch
to the ON state.
2. Create rule: Select Rule in the bottom menu. Select Source in the right
4.Advanced User Guidebook
Description
Start operation
Stop operation
Signal tested corresponds with the rule
Signal tested not correspond with the rule
Beep when it satisfies the rule
Stop once satisfying the rule
Control the display status of info frame
Select the source as CH1 or CH2
Change the Horizontal tolerance value by clicking
or
Change the Vertical tolerance value by clicking
or
Use the rule set as testing rule
Select any one from Rule1 - Rule8 as your rule name