Source Key and Programming Commands
Open/Short CAL
4.6 Open/Short CAL
Figure 4-3
Open Calibration Form
Figure 4-4
Short Calibration Form
Pressing the immediate action Open/Short CAL key directs you through performing an Open/Short
Calibration. On pressing the Open/Short CAL softkey the Open Calibration Form (see
Calibration
Form,") is displayed and the softkey menu displays a softkey labeled 'Continue' and a softkey
labeled 'Cancel'. The form shows a diagrammatic representation of how to connect the external source to
the spectrum analyzer to perform the calibration. When you press the Continue softkey, the open
calibration sweep is taken and stored in memory.
NOTE
You will get an error if you try to start an Open/Short Cal using frequency span settings that
cross band boundaries. (See
until you change the start/stop frequencies to stay within a single band.
62